13 Adjusting and calibrating the unit
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
224

Ceph - Primary diaphragm

13.2.1.4
13.2.1.4.1
Automatic adjustment: Ceph - Primary diaphragm (standard)
NOTICE
Adjustment 1
Risk of damage to unit
It is essential that the OP XG needle phantom is removed from the bite
block holder on the device before a ceph exposure is taken; otherwise,
the combisensor could collide with the phantom.
The OP XG needle phantom is
unit.
The OP XG ceph adjustment phantom is in the pan slot on the
unit [ → 209].
The ceph sensor is plugged into the ceph slot on the unit.
The ear plug holders on the ceph arm have been moved completely
apart and swung out of the beam direction (ap).
1. In the structure tree, under
Primary diaphragm"
 The
area.
2. Make SIDEXIS XG ready for exposure. [ → 206]
3. Take an exposure (64 kV/16 mA; 6.1 s) [ → 206].
 SIDEXIS XG calculates the adjustment values S1, S2, S3, and
S4 automatically from the exposure and enters them in the text
boxes of the
 The exposure is displayed in the exposure window in SIDEXIS
XG.
Installation Manual ORTHOPHOS XG 3D / Ceph
not
"2D Adjustment"
element (S010.3).
"Ceph - Primary diaphragm"
"Ceph - Primary diaphragm"
Sirona Dental Systems GmbH
in the pan bite block holder on the
"Ceph -
, click on the
menu is displayed in the action
menu.
63 03 452 D3352
D3352.031.05.08.02 10.2015