Gb/S And 10.71 Gb/S Optical Output Checks (J7230B And J7232A Opt 108 Only) - Agilent Technologies OmniBER OTN Series Performance Verification Manual

Communications performance analyzer
Hide thumbs Also See for OmniBER OTN Series:
Table of Contents

Advertisement

Verification.book Page 40 Monday, January 19, 2004 1:40 PM
1
Performance Tests
9.95 Gb/s and 10.71 Gb/s Optical Output Checks (J7230B and J7232A
Opt 108 Only)
40
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
OC-192/STM-64 Power Level and Eye Mask Tests
1 Press <Menu>, choose System > Stored Settings then press
<Select>. Select Default then Recall.
2 Press <Menu>, choose Tx/Rx > Transmitter Settings >
Physical then press <Select>. Set the Signal Mode to SONET
and Signal Rate to OC192. Ensure the Laser is set to OFF.
3 Connect the 9.95/10.71 Gb/s Optical Out port to the
oscilloscope using an Optical Patch cord. Also connect the
connect the TX Eye Clock (9.95/10.71 Gb/s) to the
oscilloscope Trigger input as shown.