Adjustment
A5 Reference Oven Adjustment (Opt 1D5 only)
A5 Reference Oven Adjustment (Opt 1D5 only)
The purpose of this procedure is to adjust the 10MHz high stability reference oscillator
(Option 1D5) frequency.
Required Equipment
Description
Frequency Counter
Frequency Standard
BNC(m)-BNC(m) Cable, 61 cm
BNC(m)-BNC(m) Cable,122cm
Procedure
Step 1. Run the adjustment program.
Step 2. Choose the A5 Adjustment(Opt.1D5) in the Spot Adjustment.
It is possible to perform this item with the All Adjustment too.
Step 3. Connect the instrument as shown in Figure 3-16.
142
Recommended Model
53181A Opt.010 and 030,
53131A Opt.010 and 030 or
53132A Opt.010 and 030
5071A or 5061B
p/n 8120-1839
p/n 8120-1840