10-24
Triggering

Remote trigger example

Table 10-2 summarizes the command sequence for basic trigger operation. These commands
set up the SourceMeter as follows:
After the unit is set up, :INIT is sent to take the unit out of idle. *TRG is sent to trigger the
unit, after which it cycles 10 times through the trigger layer. A second *TRG is required to trig-
ger the unit the second time, and it then completes the second cycle through the trigger layer.
NOTE
Table 10-2
Remote triggering example
Command
*RST
:SOUR:VOLT 10
:ARM:SOUR BUS
:ARM:COUN 2
:TRIG:DEL 0.1
:TRIG:COUN 10
:TRIG:OUTP SOUR,SENS
:TRIG:OLIN 1
:OUTP ON
:INIT
*TRG
*TRG
:OUTP OFF
:FETC?
Arm layer source: bus
Arm layer count: 2
Trigger layer delay: 0.1s
Trigger layer count: 10
Trigger layer output events: source and sense
Trigger layout trigger link output line: 1
You must allow sufficient time between the first and second *TRG commands, or the
second trigger will be ignored.
Description
Restore GPIB defaults.
Source 10V.
Select bus arm layer source (*TRG command).
Arm layer count = 2.
0.1s trigger layer delay.
Trigger layer count = 10.
Source, sense output trigger events.
Trigger output line = #1.
Turn on output.
Take unit out of idle.
Trigger first sequence.
Trigger second sequence.
Turn off output.
Request readings.