Ren (Remote Enable) - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

REN (remote enable)

The remote enable command is sent to the SourceMeter by the controller to set up the instru-
ment for remote operation. Generally, the instrument should be placed in the remote mode
before you attempt to program it over the bus. Setting REN true does not place the instrument
in the remote state. You must address the instrument to listen after setting REN true before it
goes into remote.
The SourceMeter must be in remote in order to use the following commands to trigger and
acquire readings:
:INITiate and then :FETCh?
:READ?
:MEASure?
IFC (interface clear)
The IFC command is sent by the controller to place the SourceMeter in the local, talker, lis-
tener idle states. The unit responds to the IFC command by cancelling front panel TALK or
LSTN lights, if the instrument was previously placed in one of these states.
Note that this command does not affect the status of the instrument. Settings, data, and event
registers are not changed.
With auto output off enabled (:SOURce1:CLEar:AUTO ON), the output will remain on if
operation is terminated before the output has a chance to automatically turn off.
To send the IFC command, the controller need only set the IFC line true for a minimum of
100µs.
LLO (local lockout)
Use the LLO command to prevent local operation of the instrument. After the unit receives
LLO, all of its front panel controls except OUTPUT OFF are inoperative. In this state, pressing
LOCAL will not restore control to the front panel. The GTL command restores control to the
front panel. Cycling power will also cancel local lockout.
GTL (go to local)
Use the GTL command to put a remote-mode instrument into local mode. The GTL com-
mand also restores front panel key operation.
Remote Operations
13-7

Advertisement

Table of Contents
loading

Table of Contents