17-14
SCPI Command Reference
Table 17-6 (cont.)
SOURce command summary
Command
:SOURce[1]
:CURRent
:STARt?
:STOP <n>
:STOP?
:STEP <n>
:STEP?
:SPAN <n>
:SPAN?
:CENTer <n>
:CENTer?
:VOLTage
:MODE <n>
:MODE?
:RANGe <n>|UP|DOWN|
AUTO <b>
:AUTO?
:RANGe?
[:LEVel]
[:IMMediate]
[:AMPLitude] <n>
[:AMPLitude]?
:TRIGgered
[:AMPLitude] <n>
[:AMPLitude]?
:SFACtor <n>
:STATe <b>
:STATe?
:SFACtor?
:PROTection
[:LEVel] <NRf>
[:LEVel]?
:TRIPped?
1
If source V is active.
Description
Query start level for current sweep.
Specify stop level for I-sweep; -105e-3 to
105e-3.
Query stop level for current sweep.
Specify step value for I-sweep; -210e-3 to
210e-3.
Query step value for current sweep.
Specify span; -210e-3 to 210e-3.
Query span.
Specify center point; -210e-3 to 210e-3.
Query center point.
Path to configure V-Source:
Select V-Source mode (FIXed, SWEep, or
LIST).
Query V-Source mode.
Select fixed V-Source range; 0 to 210.
Enable or disable autoranging.
Query state of autoranging.
Query V-Source range setting.
Set V-Source level (in volts):
Set specified level immediately:
Specify voltage level; -210 to 210.
Query voltage level.
Set specified level when triggered:
Specify voltage level; -210 to 210.
Query voltage level.
Set voltage scaling factor (-999.9999e+18
to +999.9999e+18).
Enable/disable voltage scaling factor.
Query voltage scaling factor state.
Query voltage scaling factor.
Path to limit output voltage:
Specify voltage limit level; -210 to 210.
Query voltage limit.
Voltage limit detected: 1 (yes), 0 (no).
Default
Source
parameter SCPI
memory
0
0
0
0
FIXed
1
21
1
ON
1
0
0
1
0
1
OFF
210

Advertisement