:CONTrol <name>
:TRACe:FEED:CONTrol <name>
Parameters
Query
Description

Select timestamp format

:FORMat <name>
:TRACe:TSTamp:FORMat <name>
Parameters
Query
Description
<name> =
NEXT
NEVer
:CONTrol?
This command is used to select the buffer control. When NEXT is selected,
the asterisk (*) annunciator turns on to indicate that the buffer is enabled.
The storage process starts when SourceMeter is taken out of idle to perform
source-measure operations.
After the buffer stores the specified number of reading arrays (as set by the
:POINTs command), the asterisk annuciator turns off to indicate that stor-
age is done.
With NEVer selected, storage into the buffer is disabled.
<name> =
ABSolute
DELTa
:FORMat?
This command is used to select the timestamp format for buffer readings.
With ABSolute selected, each timestamp is referenced to the first reading
stored in the buffer. With DELTa selected, timestamps provide the time
between each buffer reading.
SCPI Command Reference 17-101
Start or stop buffer
Fills buffer and stops
Disables buffer storage
Query buffer control
Select timestamp format
Reference to first buffer reading
Time between buffer readings
Query timestamp format