Basic Source-Measure Operation
Source-measure programming example
Table 3-7 summarizes the command sequence for a basic source-measure procedure. Note
that the steps correspond to those listed previously in Front panel source-measure procedure.
These commands set up the SourceMeter as follows:
T a ble 3-7
Basic source-measure command sequence
1
Step
1
2
3
4
5
6
1
Steps correspond to front panel steps listed previously in Front panel source-measure procedure.
2
Commands must be sent in order given.
3
Instrument must be addressed to talk after :READ? to acquire data.
Source function: volts
Source mode: fixed
Source range: 20V
Source output level: 10V
Current compliance: 10mA
Measure function: current
Measure range: 10mA
Action
Commands
*RST
Select source function,
:SOUR:FUNC VOLT
mode.
:SOUR:VOLT:MODE FIXED
Set source range,
:SOUR:VOLT:RANG 20
level, compliance.
:SOUR:VOLT:LEV 10
:SENS:CURR:PROT 10E-3
Set measure function,
:SENS:FUNC "CURR"
range.
:SENS:CURR:RANG 10E-3
Turn on output.
:OUTP ON
Read data.
:READ?
Turn off output.
:OUTP OFF
2,3
Comments
Restore GPIB defaults.
Select voltage source.
Fixed voltage source mode.
Select 20V source range.
Source output = 10V.
10mA compliance.
Current measure function.
10mA measure range.
Output on before measuring.
Trigger, acquire reading.