RS-232 interface

:LOCal
:SYSTem:LOCal
Description
:REMote
:SYSTem:REMote
Description
:RWLock <b>
:SYSTem:RWLock <b>
Parameters
Query
Description
Query timestamp
:TIME?
:SYSTem:TIME?
Query
Description
Normally, during RS-232 communications, front panel keys are operational.
However, the user may wish to lock out front panel keys during RS-232
communications. See :RWLock.
This action command is used to remove the SourceMeter from the remote
state and enables the operation of front panel keys. Note that this command
can only be sent over the RS-232 interface.
This action command is used to place the SourceMeter in the remote state.
In remote, the front panel keys will be locked out if local lockout is asserted.
See :RWLock. Note that this command can only be sent over the RS-232
interface.
<b> =
0 or OFF
1 or ON
:RWLock
This command is used to enable or disable local lockout. When enabled, the
front panel keys are locked out (not operational) when the instrument is in
remote. (See :REMote.) When disabled, the front panel keys are operational
in remote.
Removing the instrument from remote (:LOCal) restores front panel keys
operation but does not change the status of the :RWLock command.
Note that this command can only be sent over the RS-232 interface.
:TIME?
This query returns the current timestamp value.
SCPI Command Reference
Take SourceMeter out of remote
Place the SourceMeter in remote
Disable or enable front panel keys
Disable local lockout
Enable local lockout
Query state of local lockout
Query timestamp
Query timestamp