18-18
Performance Verification

Remote PreAmp output voltage accuracy

Follow the steps below to verify that Model 6430 Remote PreAmp output voltage accuracy
is within specified limits. This test involves setting the output voltage to each full-range value
and measuring the voltages with a precision digital multimeter.
1.
NOTE
2.
3.
4.
Figure 18-5
Connections for
Remote PreAmp
voltage verification
tests
With the power off, connect the digital multimeter to the Remote PreAmp IN/OUT
HIGH jack as shown in Figure 18-5. (Connect the IN/OUT HIGH jack to the DMM
INPUT HI and LO jacks using the adapters and coax cable as shown.)
Use the type of triax-to-BNC adapter that connects triax shell to BNC shell. Also be
sure that the cable shield is connected to DMM INPUT LO.
Select the SourceMeter 10µA measurement range.
Select the multimeter DC volts measuring function.
Press the Model 6430 SOURCE V key to source voltage, and make sure the source out-
put is turned on.
Model 6430
WARNING:
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
5V
HI
PK
MADE IN
V, Ω,
U.S.A.
GUARD
250V
5V
250V
5V
PEAK
PEAK
PEAK
PEAK
GUARD
SENSE
LO
4-WIRE
INPUT/
42V
SENSE
OUTPUT
PEAK
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
TRIGGER
LINK
RS232
CAUTION:
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Mainframe/PreAmp
Cable
Digital Multimeter
REMOTE
!
PreAmp
LINE FUSE
SLOWBLOW
2.5A, 250V
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
INTERLOCK-
DIGITAL I/O
IN/OUT HIGH Jack
Remote PreAmp
Input HI
Triax-to-BNC
Adapter
Coax Cable
Dual
Banana-to-BNC
Adapter
Input LO