The guard test lead IR drop is compensated for by connecting GUARD SENSE as shown in
Figure 5-14B. Sensing allows the guard voltage to be sensed (measured) at the resistor network
for better guard voltage regulation. If the remotely sensed guard voltage is less than the output
voltage of the SourceMeter, the guard voltage will be increased until the sensed guard voltage
equals the output HI voltage.
Note that in order to ensure that ohms guard current (I
50mA, output voltage from the SourceMeter must not exceed 5V (50mA × 100Ω = 5V).
NOTE
Figure 5-14
In-circuit ohms
measurements
using guard sense
A. Local Guard Sense
B. Remote Guard Sense
For 6-wire ohms guard measurements, configure the output-off state to the GUARD
mode. For details on the GUARD output-off state, see Section 12, "Output
configuration."
6430
Guard
x1
In/Out HI
I-Source V-Meter
In/Out LO
Ohms Guard Selected
6430
-
Guard
x1
+
In/Out HI
I-Source V-Meter
In/Out LO
Ohms Guard Selected
G
V,Ω GUARD
(Ohms Guard)
IN/OUT HIGH
HI (In/Out)
Guard (Cable)
LO (In/Out)
REMOTE
20kΩ
PreAmp
GUARD SENSE
V,Ω GUARD
(Ohms Guard)
IN/OUT HIGH
HI (In/Out)
Guard (Cable)
LO (In/Out)
R
REMOTE
PreAmp
20kΩ
Source-Measure Concepts
) in Figure 5-14 does not exceed
Test Lead Resistance
I
R
TL
G
1Ω
I
L
R
2
10kΩ
R
1
R
3
100kΩ
I
R
G
TL
1Ω
I
L
R
2
10kΩ
1
R
3
100kΩ
5-25