18-8
Performance Verification

Setting the source range and output value

Before testing each verification point, you must properly set the source range and output
value as outlined below.
1.
2.
3.
4.
5.
Setting the measurement range
When simultaneously sourcing and measuring either voltage or current, the measure range is
coupled to the source range, and you cannot independently control the measure range. Thus, it
is not necessary for you to set the measure range when testing voltage or current measurement
accuracy.
Compliance considerations
Compliance limits
When sourcing voltage, you can set the unit to limit current from 1fA (with PreAmp) to
105mA. Conversely, when sourcing current, you can set the unit to limit voltage from 0.2mV to
210V. The instrument output will not exceed the programmed compliance limit.
Types of compliance
There are two types of compliance that can occur: "real" and "range." Depending on which
value is lower, the output will clamp at either the displayed compliance setting ("real") or at the
maximum measurement range reading ("range").
Press either the SOURCE V or SOURCE I key to select the appropriate source
function.
Press the EDIT key as required to select the source display field. Note that the cursor
will flash in the source field while its value is being edited.
With the cursor in the source display field flashing, set the source range to the lowest
possible range for the value to be sourced using the RANGE
you should use the 20V source range to output a 20V source value. With a 20V source
value and the 20V range selected, the source field display will appear as follows:
Vsrc:+20.0000 V
With the source field cursor flashing, set the source output to the required value using
either:
The SOURCE
and
and
The numeric keys.
Note that the source output value will be updated immediately; you need not press
ENTER when setting the source value.
and
keys.
or
key. For example,