Table 5-2
Compliance examples
Display message
Cmpl: 150.000 V
Cmpl: 150.000 V
Cmpl: 150.000 V
Cmpl: 075.000 mA
Cmpl: 075.000 mA
Cmpl: 075.000 mA
Over the bus, use the appropriate SCPI commands to determine the measurement range and
the compliance setting. Once those parameters are known, compare them as previously
explained to determine the compliance in effect.
When sourcing current, use the following commands to acquire the measurement range and
the compliance setting:
VOLTage:RANGe?
VOLTage:PROTection?
When sourcing voltage, use the following commands to acquire the measurement range and
the compliance setting:
CURRent:RANGe?
CURRent:PROTection?

Overheating protection

Proper ventilation is required to keep the SourceMeter from overheating. See the
"CAUTION" located at the beginning of Section 3 for details on maintaining proper
ventilation.
The SourceMeter has an over-temperature protection circuit that will turn the output off in
the event that the instrument overheats. If the output trips due to overheating, a message indi-
cating this condition will be displayed. You will not be able to turn the output back on until the
instrument cools down.
Compliance setting
Setting
150V
150V
150V
75mA
75mA
75mA
Source-Measure Concepts
Measurement range
Display message Range
---.---V
200V
--.----V
20V
---.---mV
200mV
---.---mA
100mA
--.---mA
10mA
-.-----mA
1mA
Query voltage measurement range.
Query voltage compliance limit.
Query current measurement range.
Query current compliance limit.
Actual compliance
Value
Type
150V
Real
21V
Range
210mV
Range
75mA
Real
10.5mA
Range
1.05mA
Range
5-5