14-18
Status Structure
Table 14-7
Event enable registers commands
Command
*ESE <NDN> or <NRf>
*ESE?
STATus
:OPERation
:ENABle <NDN> or <NRf>
:ENABle?
:MEASurement
:ENABle <NDN> or <NRf>
:ENABle?
:QUEStionable
:ENABle <NDN> or <NRf>
:ENABle?
Parameters:
<NDN> = #Bxx...x
= #Hx
= #Qx
<NRf>
= 0 to 65535
Note: Power-up and STATus:PRESet resets all bits of all enable registers to 0. *CLS has no effect.
Programming example - program and read register set
The command sequence in Table 14-8 programs and reads the measurement register set.
Registers are read using the binary format (which directly indicates which bits are set). The
command to select format (FORMat:SREGister) is documented in Table 14-2.
Table 14-8
Program and read register programming example
Command
FORM:SREG BIN
STAT:MEAS:ENAB 512
STAT:MEAS:COND?
STAT:MEAS?
Description
Program Standard Event Enable Register. See Parameters.
Read Standard Event Enable Register.
STATus Subsystem:
Operation Event Enable Register:
Program enable register. See Parameters.
Read enable register.
Measurement Event Enable Register:
Program enable register. See Parameters.
Read enable register.
Questionable Event Enable Register:
Program enable register. See Parameters.
Read Measurement Event Enable Register:
Binary format (each x = 1 or 0)
Hexadecimal format (x = 0 to FFFF)
Octal format (x = 0 to 177777)
Decimal format
Description
Select binary format to read registers.
Enable BFL (buffer full).
Read Measurement Condition Register.
Read Measurement Event Register.
Default
Note

Advertisement