12-10
Digital I/O Port, Interlock, and Output Configuration
Output configuration programming example
Table 12-3 lists the command sequence for output configuration. These commands set up the
SourceMeter as follows:
NOTE
Table 12-3
Output configuration programming example
Command
*RST
:SOUR:VOLT 10
:OUTP:INT:STAT ON
:OUTP:SMOD NORM
:SOUR:CLE:AUTO ON
:READ?
*Connect pins 8 and 9 of Digital I/O Port to simulate closed interlock switch.
Interlock: enabled
Output-off mode: normal
Auto-off mode: on
Connect pins 8 and 9 of the Digital I/O Port together to simulate a closed interlock
switch. Otherwise, the unit will not turn on its output when the measurement is made.
Description
Restore GPIB defaults.
Output 10V.
Enable interlock.*
Select normal output-off mode.
Enable auto-off mode.
Trigger and acquire readings.