Figure 19-3 Voltage Burden Calibration Connections - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

19-16
Calibration
Figure 19-3
Voltage burden
calibration
connections
7.
8.
9.
Model 6430
WARNING:
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
HI
5V
PK
MADE IN
V, Ω,
U.S.A.
GUARD
5V
250V
5V
250V
PEAK
PEAK
PEAK
PEAK
GUARD
SENSE
LO
4-WIRE
INPUT/
42V
SENSE
OUTPUT
PEAK
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
TRIGGER
LINK
RS232
CAUTION:
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Mainframe/PreAmp
Cable
Digital Multimeter
Again, adjust the display to agree with the DMM reading.
Press ENTER. The instrument will prompt you for the third DMM reading:
DMM RDG: +000.0000mV
Use
, ▲, ▼,
, ENTER or EXIT.
Adjust the display to agree with the DMM reading, then press ENTER. The unit will
return to the calibration menu.
REMOTE
!
PreAmp
LINE FUSE
SLOWBLOW
2.5A, 250V
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
INTERLOCK-
DIGITAL I/O
IN/OUT
HIGH Jack
Remote PreAmp
Input HI
Triax-to-BNC
Adapter
Coax Cable
Dual
Banana-to-BNC
Adapter
Input LO