:ADVanced:NTOLerance <NRf>
[:SENSe[1]]:AVERage:ADVanced:NTOLerance <NRf>
Parameters
<NRf> =
Query
:NTOLerance?
Description
When the advanced filter is enabled, using the next command, a noise win-
dow is used with the moving filter. This command is used to specify the
noise window. If readings are within the noise window, the moving filter
operates normally. If, however, a reading falls outside the window, the
buffer is flushed of old readings and filled with the new reading.
:ADVanced[:STATe] <b>
[:SENSe[1]]:AVERage:ADVanced[:STATe] <b>
Parameters
<b> =
Query
[:STATe]?
Description
This command is used to enable or disable the advanced filter. When
enabled, the noise window is used with the moving filter. When disabled,
the noise window is not used.
Set moving filter noise filter
0 to 105
Specify filter noise filter in %
Query filter noise filter value
Enable/disable advanced filter
0 or OFF
Disable advanced filter
1 or ON
Enable advanced filter
Query state of advanced filter
SCPI Command Reference