Figure 10-6 Trigger Link Connections - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

10-10
Triggering
The Trigger Link connections for this test system are shown in Figure 10-6. Trigger Link of
the SourceMeter is connected to Trigger Link (IN or OUT) of the switching mainframe. Note
that with the default trigger settings of the switching mainframe, line #1 is an input, and line #2
is an output.
Figure 10-6
Trigger link
connections
For this example, the SourceMeter and Switching Mainframe are configured as follows.
SourceMeter setup
Step 1: Restore bench defaults
Press the MENU key, select SAVESETUP, then press the ENTER key. From the
SAVESETUP menu, select GLOBAL, then press ENTER. From the GLOBAL SETUP
MENU, select RESET, then press ENTER. Select BENCH, then press ENTER.
Step 2: Set up trigger parameters
Press the CONFIG key, and then the TRIG key to access the CONFIGURE TRIGGER
menu. Select TRIG-LAYER, and then press ENTER to access the CONFIGURE TRIG-
LAYER menu.
Step 3: Set trigger-in event to TRIGGER-LINK
Select TRIGGER-IN, then press ENTER. Select TRIGGER-LINK, then press ENTER.
Step 4: Set trigger input line to #2
Select #2, then press ENTER three times to return to CONFIGURE TRIG-LAYER menu.
Step 5: Set trigger output line to # 1
Select TRIGGER-OUT, then press ENTER. Select LINE, then press ENTER. Select #1,
then press ENTER.
7001 or 7002 Switch System
WARNING:
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
MADE IN USA
IN
OUT
CAUTION:
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Trigger
Link
6430 SourceMeter
WARNING:
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
5V
HI
PK
V, Ω
GUARD
5V
250V
5V
250V
PEAK
PEAK
PEAK
PEAK
GUARD
SENSE
LO
4-WIRE
INPUT/
42V
SENSE
OUTPUT
PEAK
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
RS232
CAUTION:
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Trigger
Link Cable
(8501)
MADE IN
U.S.A.
REMOTE
!
PreAmp
LINE FUSE
SLOWBLOW
2.5A, 250V
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
INTERLOCK-
TRIGGER
DIGITAL I/O
LINK
Trigger
Link

Advertisement

Table of Contents
loading

Table of Contents