17-70
SCPI Command Reference

Set delay

:DELay <n>
:SOURce[1]:DELay <n>
Parameters
Query
Description
:AUTO <b>
:SOURce[1]:DELay:AUTO <b>
Parameters
Query
Description
<n> =
0 to 999.9999
MINimum
MAXimum
DEFault
:DELay?
:DELay? DEFault
:DELay? MINimum
:DELay? MAXimum
This command is used to manually set a delay (settling time) for the source.
After the programmed source is turned on, this delay occurs to allow the
source level to settle before a measurement is taken. Note that this delay is
the same for both the I-Source and V-Source.
Do not confuse this source delay with the trigger delay. The source delay is
part of the device action (SDM cycle) while the trigger delay occurs before
the device action. See Section 10, Trigger model for more information.
Auto delay can instead be used to automatically set the source delay. See the
next command.
<b> =
0 or OFF
1 or ON
:AUTO?
This command is used to enable or disable auto delay. When enabled, the
instrument will automatically select a delay period that is appropriate for the
present source/measure setup configuration. See Table 3-2. *RST and
SYST:PRES default is OFF.
Manually set source delay
Specify delay in seconds
0 seconds
999.9999 seconds
0.003
Query delay
Query *RST default delay
Query lowest allowable delay
Query highest allowable delay
Enable/disable auto delay for source
Disable auto delay
Enable auto delay
Query state of auto delay