Select Function Mode - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

Select function mode

[:MODE] <name>
:SOURce[1]:FUNCtion[:MODE] <name>
Parameters
Query
Description
Select sourcing mode
:MODE <name>
:SOURce[1]:CURRent:MODE <name>
:SOURce[1]:VOLTage:MODE <name>
Parameters
Query
Description
NOTE
The sourcing mode will default to FIXed whenever the SourceMeter goes to the local
state.
<name> =
VOLTage
CURRent
MEMory
[:MODE]?
This command is used to select the source mode. With VOLTage selected,
the V-Source will be used, and with CURRent selected, the I-Source will be
used.
With MEMory selected, a memory sweep can be performed. Operating set-
ups (up to 100) saved in memory can be sequentially recalled. This allows
multiple source/measure functions to be used in a sweep.
<name> =
FIXed
LIST
SWEep
:MODE?
This command is used to select the DC sourcing mode for the specified
source. The three modes are explained as follows:
FIXed — In this DC sourcing mode, the specified source will output a fixed
level. Use the :RANGe and :AMPLitude commands to specify the fixed
source level. See Select range and Set amplitude for fixed source.
LIST — In this mode, the source will output levels that are specified in a
list. See Configure list for commands to define and control the execution of
the list.
SWEep — In this mode, the source will perform a voltage, current or mem-
ory sweep. See Configure voltage and current sweeps and Configure mem-
ory sweep for commands to define the sweep.
SCPI Command Reference
Select source mode
Select voltage mode
Select current mode
Select memory mode
Query selected source
Select DC sourcing mode for I-Source
Select DC sourcing mode for V-Source
Select fixed sourcing mode
Select list sourcing mode
Select sweep sourcing mode
Query DC sourcing mode
17-65