Digital Output Configuration - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

EOT/BUSY line
Line 4 can be used for a normal bit pattern, EOT (End-of-Test), or BUSY signal, depending
on the selected END OF TEST mode.
NOTE
+5V output
The Digital I/O Port provides a +5V output that can be used to drive external logic circuitry.
Maximum current output for this line is 300mA. This line is protected by a self-resetting fuse
(one hour recovery time).
Digital output configuration
There are two basic methods to connect external components to the digital output lines, sink
operation and source operation.
Sink operation
Figure 12-2 shows the basic output configuration for sink operation. Note that the external
relay coil is connected between the digital output line (pins 1 to 4) and +5V (pin 7). With this
configuration, the digital output line must be set LO to energize the relay, and the maximum
sink current is 500mA.
Figure 12-2
Sink operation
+5V
Digital I/O Port, Interlock, and Output Configuration
See Section 11 for details on performing limit tests and Section 10 for information
on programming the SourceMeter to respond to the SOT (start-of-test) pulse from a
handler.
SourceMeter
Pin 7
Pins 1-4
Pin 9
Digital I/O
Port
External
Relay
Maximum sink current: 500mA
12-3
To other
Circuits

Advertisement

Table of Contents
loading

Table of Contents