OUTPut <event list>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OUTPut <event list>
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut <event list>
Parameters
Arm Layer Triggers
<event list >: TENTer
Trigger Layer Triggers
<event list>: SOURce
NOTE
Each event in the list must be separated by a comma (i.e., :arm:output source, delay,
sense).
Query
:OUTPut?
Description
This command is used to specify when trigger pulses occur on the specified
output trigger line of the Trigger Link (:OLINe).
Arm Layer Triggers — With TEXit selected, an output trigger will occur
when exiting the trigger layer. With TENTer selected, an output trigger will
occur when entering the trigger layer. With NONE selected, the arm layer
output trigger is disabled.
Trigger Layer Triggers — You can specify from one to all three events. Each
event in the list must be separated by a comma (,).
The SOURce, DELay and MEASure events refer to the Source-Delay-
Measure (SDM) cycle. This is the Device Action in the Trigger Model. See
Figures 10-7 and 10-8. With SOURce specified, an output trigger occurs
after the source is set. With DELay specified, an output trigger occurs after
the delay period. With MEASure specified, an output trigger occurs after the
measurement.
NOTE
When disabling triggers, the NONE parameter must be sent alone (i.e., trig:outp
none). If it is listed with any of the other event parameters, NONE will be ignored.
Trigger on entering trigger layer
TEXit
Trigger on exiting trigger layer
NONE
Disable arm layer output trigger
Output trigger after source level is set
DELay
Output trigger after delay period
SENSe
Output Trigger after measurement
NONE
Disable trigger layer triggers
Query output trigger event(s)
SCPI Command Reference 17-107