Table 9-6 Custom Sweep Programming Example; Table 9-7 Source Memory Sweep Commands - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

Custom sweep programming example
As an example of custom sweep operation, assume a five-point sweep with the following
parameters:
Source Function: volts
Sense Function: current
Voltage Sweep Mode: list (custom sweep)
Sweep Voltage Points: 7V, 1V, 3V, 8V, 2V
Current Compliance: 100mA
Source Delay: 100ms
Table 9-6 summarizes the basic remote command sequence for performing the custom
sweep described above.
Table 9-6
Custom sweep programming example
Command
*RST
:SENS:FUNC:CONC OFF
:SOUR:FUNC VOLT
:SENS:FUNC 'CURR:DC'
:SENS:CURR:PROT 0.1
:SOUR:VOLT:MODE LIST
:SOUR:LIST:VOLT 7,1,3,8,2
:TRIG:COUN 5
:SOUR:DEL 0.1
:OUTP ON
:READ?
Source memory sweep commands
Table 9-7 summarizes remote commands used for custom sweep operation. See Section 17,
Configure memory sweep, for more details on these commands.
Table 9-7
Source memory sweep commands
Command
:SOURce:FUNCtion MEM
:SOURce:MEMory:POINts <n>
:SOURce:MEMory:STARt <n>
:SOURce:MEMory:RECall <n>
:SOURce:SAVE <n>
Description
Restore GPIB default conditions.
Turn off concurrent functions.
Volts source function.
Current sense function.
100mA current compliance.
List volts sweep mode.
7V, 1V, 3V, 8V, 2V sweep points.
Trigger count = # sweep points.
100ms source delay.
Turn on source output.
Trigger sweep, request data.
Description
Select memory sweep mode.
Specify number of sweep points (n = points).
Select source memory start location (n =location).
Return to specified setup (n = memory location).
Save setup in memory (n = memory location).
Sweep Operation
9-21

Advertisement

Table of Contents
loading