Keithley 6430 Instruction Manual page 478

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

A-2
Specifications
SOURCE SPECIFICATIONS
Voltage Programming Accuracy (4-wire sense)
Programming
Range
Resolution
200.000 mV
5 µV
2.00000
V
50 µV
20.0000
V
500 µV
200.000
V
5 mV
TEMPERATURE COEFFICIENT (0°–18°C & 28°–40°C): ±(0.15 × accuracy specification)/°C.
MAX. OUTPUT POWER: 2.2W (four quadrant source or sink operation).
SOURCE/SINK LIMITS: ±21V @ ±105mA, ±210V @ ±10.5mA.
VOLTAGE REGULATION: Line: 0.01% of range. Load: 0.01% of range + 100µV.
NOISE 10Hz–1MHz (p-p): 10mV.
OVER VOLTAGE PROTECTION: User selectable values, 5% tolerance. Factory default = None.
CURRENT LIMIT: Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
Current Programming Accuracy (with remote preamp)
Programming
Range
Resolution
1.00000
pA
50 aA
10.0000
pA
500 aA
100.000
pA
5 fA
1.00000
nA
50 fA
10.0000
nA
500 fA
100.000
nA
5 pA
1.00000
µA
50 pA
10.0000
µA
500 pA
100.000
µA
5 nA
1.00000 mA
50 nA
10.0000 mA
500 nA
100.000 mA
5 µA
Current Programming Accuracy (without remote preamp)
Programming
Range
Resolution
1.00000
µA
50 pA
10.0000
µA
500 pA
100.000
µA
5 nA
1.00000 mA
50 nA
10.0000 mA
500 nA
100.000 mA
5 µA
TEMPERATURE COEFFICIENT (0°–18°C & 28°–40°C): ±(0.15 × accuracy specification)/°C.
MAX. OUTPUT POWER: 2.2W (four quadrant source or sink operation).
SOURCE/SINK LIMITS: ±10.5mA @ 210V, ±105mA @ 21V.
CURRENT REGULATION: Line: 0.01% of range. Load: 0.01% of range + 1fA.
VOLTAGE LIMIT: Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
For sink mode, 1pA to 100mA range, accuracy is ±(0.15% + offset * 4).
1
Voltage source accuracies are not affected by the remote preamp.
2
1
2
Accuracy (1 Year)
Noise
23°C ±5°C
(peak-peak)
±% rdg. + volts
0.1Hz – 10Hz
0.02% + 600 µV
5 µV
0.02% + 600 µV
50 µV
0.02% +
2.4 mV
500 µV
0.02% +
24 mV
5 mV
Accuracy (1 Year)
Noise
1
23°C ±5°C
(peak-peak)
±% rdg. +amps
0.1Hz – 10Hz
1.0
% +
10 fA
5 fA
0.50
% +
30 fA
10 fA
0.15
% +
40 fA
20 fA
0.050 % + 200 f A
50 fA
0.050 % +
2 pA
500 fA
0.050 % +
20 pA
3 pA
0.050 % + 300 pA
20 pA
0.050 % +
2 nA
200 pA
0.031 % +
20 nA
500 pA
0.034 % + 200 nA
5 nA
0.045 % +
2 µA
50 nA
0.066 % +
20 µA
500 nA
Accuracy (1 Year)
Noise
1
23°C ±5°C
(peak-peak)
±% rdg. +amps
0.1Hz – 10Hz
0.035 % + 600 pA
20 pA
0.033 % +
2 nA
200 pA
0.031 % +
20 nA
500 pA
0.034 % + 200 nA
5 nA
0.045 % +
2 µA
50 nA
0.066 % +
20 µA
50 nA

Advertisement