Section 10: Digital I/O and Triggering
6.
Choose DELAY, set the delay to 2s, then press ENTER.
7.
Press EXIT to return to normal display.
8.
Turn on the OUTPUT ON, then press TRIG. A two second delay will occur before the first
measurement. The unit will cycle through measurements indefinitely with a 1s interval
between measurements.
9.
Turn the OUTPUT off to stop taking readings.

Remote triggering

Remote trigger commands
Trigger commands are listed in
The trigger event detector remembers if an event has been detected since the last
trigger.wait call. The trigger.clear() command clears the trigger's event detector and
discards the previous history of command interface trigger events.
The trigger.wait function will wait timeout seconds for a GPIB GET command (see
Section
command interface, or a *TRG message on the command interface for all other interfaces. If one
or more of these trigger events were previously detected, this function will return immediately. After
waiting for a trigger with this function, the event detector will be automatically reset and re-armed
regardless of the number of events detected.
It is important to note the following:
A synchronization line event detector sets the trigger overrun if an edge is detected while
the event detector is in the detected state
You can use the digo.trigger[N].clear command to retest the event detector
detected state and to clear the trigger overrun indication
Review the trigger overrun to confirm that a trigger was not missed
Command
trigger.clear()
triggered = trigger.wait(timeout)
Remote trigger example
The example below clears triggers and then enables a 30 second timeout to wait for a GET or
*TRG trigger. A maximum of 50 *TRG commands can be used before the input buffer overflows.
smua.reset()
triggered = trigger.wait(30)
smua.source.output =smua.OUTPUT_ON
*TRG
reading = smua.measure.v()
10-14
Table
11) or a *TRG message (see
Return to
Series 2600 System SourceMeter® Instruments Reference Manual
10-3. See
Section 12
Appendix
C) on the GPIB interface if that is the active
Description
Clears triggers pending.
Waits the timeout seconds for GET or *TRG trigger.
- Restore 2600 defaults.
- Wait 30s for GET or *TRG.
- Turn on output.
- Trigger reading.
- Get voltage reading.
Section Topics
for more details.