17-98
SCPI Command Reference

Reset timestamp

:RESet
:SYSTem:TIME:RESet
Description
Auto reset timestamp
:RESet:AUTO <b>
:SYSTem:TIME:RESet:AUTO <b>
Parameters
Query
Description
Auto range change mode
:RCMode <name>
:SYSTem:RCMode <name>
Parameters
Query
Description
This action command is used to reset the absolute timestamp to 0 seconds.
The timestamp also resets to 0 seconds every time the SourceMeter is turned
on.
<b> =
1 or ON
0 or OFF
:AUTO?
:RES:AUTO enables or disables auto timestamp reset. When enabled, the
timestamp will be automatically reset when exiting the idle layer of the trig-
ger model. This command is intended for use with READ?/INIT when tak-
ing more than one reading.
<name> = SINGle
MULTiple
:RCMode?
This command controls the auto range change mode. In the SINGle mode,
the SourceMeter will auto range only after first taking a reading. In the
MULTiple mode, the SourceMeter will auto range up on compliance in the
Delay phase of the Source-Delay-Measure cycle, thereby minimizing the
chance of a SourceMeter being in compliance in a multiple-SourceMeter
system. A SourceMeter can downrange only once a reading has been taken.
In the MULTiple mode, you can control the soak time using the
:SOUR:SOAK command. See SOURce subsystem. Note that you can use
the LLIMIT and ULIMIT commands to control auto range limits. See
SENSe1 subsystem.
Reset timestamp
Reset timestamp when exiting idle
Enable auto timestamp reset
Disable auto timestamp reset
Query enabled/disabled auto timestamp reset
state
Control auto range change mode
Single mode
Multiple mode
Query auto range change mode