18-22
Performance Verification
1pA-100nA range accuracy
1.
Figure 18-7
Connections for
1pA-100nA
range current
verification tests
With the power off, connect the digital multimeter and calibration standard to the
Model 6430 mainframe and Remote PreAmp as shown in Figure 18-7. (Connect the
mainframe INPUT/OUTPUT HI and LO jacks to DMM INPUT HI and LO respec-
tively. Connect the PreAmp IN/OUT HIGH jack directly to the Model 5156 OUTPUT
jack, and connect the BNC shorting cap to the 100GΩ resistance jack. Also be sure to
remove the link between the Model 5156 SHIELD and CHASSIS jacks.)
Model 6430
WARNING:
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
5V
HI
PK
MADE IN
V, Ω,
U.S.A.
GUARD
5V
250V
5V
250V
PEAK
PEAK
PEAK
PEAK
GUARD
SENSE
LO
4-WIRE
INPUT/
42V
SENSE
OUTPUT
PEAK
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
TRIGGER
LINK
RS232
CAUTION:
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Digital Multimeter
Connect PreAmp IN/OUT HIGH Jack
REMOTE
!
PreAmp
Directly to 5156 OUTPUT Jack
using triax male-to-male Adapter
LINE FUSE
SLOWBLOW
2.5A, 250V
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
INTERLOCK-
DIGITAL I/O
Mainframe/PreAmp Cable
100GΩ
1 nF
100nF
Model 5156 Electrometer Calibration Standard
Input HI
Input LO
BNC Shorting Plug
(Connect to Resistor
Being used)
10GΩ
1GΩ
100MΩ
Note: Remove Link Between
SHIELD and CHASSIS