Connections To Dut - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

2-6
Connections

Connections to DUT

NOTE
WARNING
CAUTION
Sensing methods
Basic source-measure operations are performed using either 2-wire sense connections
(Figure 2-2) or 4-wire sense connections (Figure 2-3). See Section 4, Ohms sensing for
additional information.
Figure 2-2
Two-wire
sense
DUT
connections
Connect to earth safety
ground using #18 AWG
wire or larger
Connection drawings in this manual are shown using the Remote PreAmp. If not
using the Remote PreAmp, make connections to the equivalent banana jack termi-
nals on the rear panel of the mainframe.
To prevent injury from electric shock, DO NOT touch the triax cables of
the Remote PreAmp while the SourceMeter is turned on or is connected to
an external source that is turned on.
When using the Remote PreAmp, DO NOT use the INPUT/OUTPUT HI
and 4-WIRE SENSE HI banana jacks on the mainframe. Using these ter-
minals while the Remote PreAmp is plugged in will result in errors and/or
noise, and possible damage to the Remote Amplifier.
Noise Shield
HI
Triax Cable
LO
Test
Fixture
WARNING Guard is at the same voltage level as
input/output high. Therefore, if a
hazardous voltage (≥42V peak) is
present on input/output high, it is also
present on the guard terminals of the
Remote PreAmp and mainframe.
*Keithley part number: CA-176-1D.
KEITHLEY
Preamp Cable*
6430
REMOTE
PreAmp
Connect to REMOTE
PreAmp connector on rear
panel of mainframe

Advertisement

Table of Contents
loading

Table of Contents