17-106 SCPI Command Reference
Description
NOTE
:ILINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:ILINe <NRf>
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe <NRf>
Parameters
Query
Description
:OLINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OLINe <NRf> Select output line; arm layer
:TRIGger[:SEQuence[1]][:TCONfigure]:OLINe <NRf>
Parameters
Query
Description
When TLINk is the selected Trigger Layer control source, and an event
detector in the Trigger Layer is enabled, operation will hold up at that detec-
tor until an input trigger is received via the Trigger Link. When the event
detector is disabled, operation will not hold up. It continues on and performs
the appropriate action.
A Trigger Layer event detector is enabled by including the parameter name
in the event list for the INPut command. For example, to enable the Source
Event Detector and Measure Event Detector, send the following command:
:TRIGger:INPut SOURce, SENSe
The Delay Event Detector will be disabled since the DELay parameter is not
included in the above event list.
To disable all the Trigger Layer event detectors, the NONE parameter must be sent
alone (i.e. trigger:input none). If it is listed with any of the other parameters, NONE
will be ignored.
<NRf> =
1
2
3
4
:ILINe?
This command is used to select input lines for the Trigger Link. For normal
operation, Trigger Link input and output (see :OLINe) should not share the
same line.
<NRf> =
1
2
3
4
:OLINe?
This command is used to select output lines for the Trigger Link. For normal
operation, Trigger Link input and output (see :ILINe) should not share the
same line.
Select input line; arm layer
Select input line; trigger layer
Line #1
Line #2
Line #3
Line #4
Query input trigger line
Select output line; trigger layer
Line #1
Line #2
Line #3
Line #4
Query output trigger line