Figure 10-1 Trigger Model (Front Panel Operation) - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

Figure 10-1
Trigger model
(front panel
operation)
Immediate
GPIB
Timer
Manual
TLink
↓Stest
↑Stest
↑↓Stest
Immediate
Trigger Link
= Bench Default
= Output Trigger
Idle
Bypass
Arm Event
Detector
?
NEVER
Arm Event
Arm-In
Detector
Event
Arm-Out Event
Bypass
Source Event
Detector
?
Never
On/Off
Source Event
Detector
Trigger Delay
SOURCE
Action
Trigger Out Event
On/Off
Delay Event
Trigger-In
Detector
Source
DELAY
Action*
Trigger Out Event
Measure
On/Off
Event
Detector
MEASURE
Action
Trigger Out Event
*
Soak time takes the place of the delay time only during the first SDM
cycle after initial sweep trigger if the unit is in the MULTIPLE mode.
See Section 6, Auto range change mode .
Turn Output ON
Once
No
Another
Yes
Arm
?
Arm-Out Event
On/Off
On/Off
Once
No
Another
Yes
Trigger
?
0.0 sec
On/Off
0.001 sec
On/Off
On/Off
Triggering
10-3
Idle
Arm
Arm
Counter
Layer
1
Trigger
Layer
Trigger
Counter
1
MEASURE
Action
CONV
CONV
Filter
Process
(Repeat)
CONV
CONV = Reading Conversion

Advertisement

Table of Contents
loading