Output Subsystem - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

17-52
SCPI Command Reference

OUTPut subsystem

This subsystem is used to control the output of the selected source, and the interlock. These
commands are summarized in Table 17-4.
Turn source on or off
[:STATe] <b>
:OUTPut[1][:STATe] <b>
Parameters
Query
Description
NOTE
Interlock control
:STATe <b>
:OUTPut[1]:INTerlock:STATe <b>
Parameters
Query
Description
<b> =
0 or OFF
1 or ON
:OUTPut?
This command is used to turn the source output on or off. Measurements
cannot be made while the source is off.
Turning the source off places the SourceMeter in the idle state. The only
exception to this is when source auto clear is enabled. In this mode, the
source turns on during each source phase of the SDM cycle and then turns
off after each measurement. See :SOURce[1]:CLEar:AUTO in SOURce
subsystem.
The :SOURce:CLEar command will also turn the source off.
<b> =
0 or OFF
1 or ON
:STATe?
This command is used to enable or disable the hardware interlock. When
enabled, the source cannot be turned on unless the interlock line (pin 8 of
the rear panel Interlock - Digital I/O connector) is pulled to a logic low
state. When the interlock line goes to a logic high state, the source turns off.
See Section 12, Digital I/O port and Safety interlock for details using inter-
lock with a test fixture.
When disabled, the logic level on the interlock line has no effect on the out-
put state of the source.
Turn source on or off
Turn source off (standby)
Turn source on (operate)
Query state of source
Control hardware interlock
Disable interlock
Enable interlock
Query state of interlock

Advertisement

Table of Contents
loading

Table of Contents