Keithley 6430 Instruction Manual page 325

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

Table 17-6 (cont.)
SOURce command summary
Command
:SOURce[1]
:VOLTage
:STARt <n>
:STARt?
:STOP <n>
:STOP?
:STEP <n>
:STEP?
:SPAN <n>
:SPAN?
:CENTer <n>
:CENTer?
:SOAK <NRf>
:SOAK?
:SWEep
:SPACing <name>
:SPACing?
:POINts <n>
:POINts?
:DIRection <name>
:DIRection?
:RANGing <name>
:RANGing?
:LIST
:CURRent <NRf>
:APPend <NRf>
:POINts?
:CURRent?
:VOLTage <NRf>
:APPend <NRf>
:POINts?
:VOLTage?
Description
Specify start level for V-sweep; -210 to 210.
Query start level for voltage sweep.
Specify stop level for V-sweep; -210 to 210.
Query stop level for voltage sweep.
Specify step value for V-sweep; -420 to 420.
Query step value for voltage sweep.
Specify span; -420 to 420.
Query span.
Specify center point; -420 to 420.
Query center point.
Set first sweep point soak time (0.00000 to
9999.999s).
Query soak time.
Configure SWEep source mode:
Select sweep spacing type (LINear or
LOGarithmic).
Query sweep spacing.
Specify number of sweep points (2 to 2500).
Query number of points in sweep.
Sweep from start to stop (UP) or from stop to
start (DOWN).
Query sweep direction.
Select source ranging mode (BEST, AUTO,
or FIXed).
Query source ranging mode.
Configure LIST source mode:
Create list of I-Source values; -0.105 to
0.105.
Add I-Source values to end of list.
Query number of source values in list.
Query I-Source list.
Create list of V-Source values; -210 to 210.
Add V-Source values to end of list.
Query number of source values in list.
Query V-Source list.
SCPI Command Reference
Default
parameter SCPI
0
0
0
0
0
0.00000
LINear
2500
UP
BEST
17-15
Source
memory

Advertisement

Table of Contents
loading

Table of Contents