Set Compliance Limit - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

Set compliance limit

[:LEVel] <n>
[:SENSe[1]]:CURRent[:DC]:PROTection[:LEVel] <n>
[:SENSe[1]]:VOLTage[:DC]:PROTection[:LEVel] <n>
Parameters
Query
Description
NOTE
You cannot set compliance less than 0.1% of the present measurement range.
:TRIPped?
[:SENSe[1]]:CURRent[:DC]:PROTection:TRIPped?
[:SENSe[1]]:VOLTage[:DC]:PROTection:TRIPped?
Description
<n> =
-105e-3 to 105e-3
-210 to 210
DEFault
MINimum
MAXimum
:LEVel?
:LEVel? DEFault
:LEVel? MINimum
:LEVel? MAXimum
This command is used to set compliance limits. A current compliance limit
is set for the V-Source, and a voltage compliance limit is set for the I-
Source. The SourceMeter cannot source levels that exceed these specified
limits.
The :SENSe:CURRent:PROTection[:LIMit] command is used to set the
current compliance for the V-Source and the
:SENSe:VOLTage:PROTection[:LIMit] command is used to set the voltage
compliance for the I-Source.
This command is used to determine if the source is in compliance. If a "1" is
returned, then the source is in compliance. A "0" indicates that the source is
not in compliance.
The :CURRent:PROTection:TRIPped? command is used to check the com-
pliance state of the V-Source, and the :VOLTage: PROTection:TRIPped?
command is used to check the compliance state of the I-Source.
SCPI Command Reference
Set current compliance
Set voltage compliance
Current compliance limit
Voltage compliance limit
105uA, 21V
-105e-3A, -210V
105e-3A, 210V
Query compliance value
Query *RST default compliance
Query minimum allowable compliance
Query maximum allowable compliance
Query current compliance state
Query voltage compliance state
17-59