Figure 2-1 Basic Input/Output Configurations - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

2-4
Connections
Remote PreAmp triax connectors
The electrical configuration of each triax connector is shown in Figure 2-1A. The center
conductor of the connector (and triax cable) is HI (input/output or sense), and the inner shield
is cable guard. The outer shield (shell) of each triax connector is input/output LO.
NOTE
The outer shells of the triax connectors are connected to Input/Output LO. Therefore, for
floating source-measure operations, a voltage potential will be present on the shells of the triax
connectors. Even with the output of the SourceMeter off, voltage could be applied from the
external test circuit. As a general rule, do not touch the triax cables while any power is present.
WARNING
Figure 2-1
Basic input/output
configurations
The 6430-322-1A triax cable (which is a supplied accessory) is terminated with a
triax connector on one end and booted alligator clips on the other end. (See Figure
2-1B.) When connected to the Remote PreAmp, the alligator clip with the red boot is
HI, the one with the black boot is GUARD and the one with the green boot is LO.
To prevent injury from electric shock, DO NOT touch the triax cables of
the Remote PreAmp while the SourceMeter is turned on or any external
source is turned on.
Center Conductor
Inner Shield
Outer Shield
A) Remote PreAmp Triax Connector Configuration
HI (center conductor)
Guard (inner shield)
LO (outer shell)
B) Terminal identification for 6430-322-1A cable
C) Mainframe Input/Output Configuration
HI (In/Out or Sense)
GUARD (Cable)
LO (In/Out)
6430 Mainframe
HI
INPUT/OUTPUT
LO
42V Peak DC Max
Zener
Clamp
10.5mA Max
Chassis Ground Screw
(on Rear Panel)
Chassis
Ground
Triax to Alligator Clip
or
Red boot
Black boot
Green boot

Advertisement