Keithley 6430 Instruction Manual page 321

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

Table 17-5 (cont.)
SENSe command summary
Command
[:SENSe[1]]
:CURRent[:DC]
:NPLCycles <n>
:NPLCycles?
:PROTection
[:LEVel] <n>
[:LEVel]?
:TRIPped?
:VOLTage[:DC]
:RANGe
[:UPPer]
<n>|UP|DOWN
[:UPPer]?
:AUTO <b>
:AUTO?
:ULIMit?
:LLIMit <n>
:LLIMit?
:NPLCycles <n>
:NPLCycles?
:PROTection
[:LEVel] <n>
[:LEVel]?
:TRIPped?
:RESistance
:MODE <name>
:MODE?
:OCOMpensated <b>
:OCOMpensated?
1
If source V is active.
2
If source I is active and auto ohms is disabled.
Description
Specify integration rate (in line cycles): 0.01
to 10.
Query integration rate.
Path to configure current compliance:
Specify current limit for V-Source; -105e-3
to 105e-3.
Query current compliance limit.
In current compliance: 1 (yes), 0 (no).
Path to configure volts:
Configure measurement range:
Select range by specifying the expected
voltage reading; 0 to ±210.
Query range.
Enable or disable auto range.
Query auto range.
Returns V compliance range.
Set auto range lower limit (-21 to 21).
Query auto range lower limit.
Specify integration rate (in line cycles): 0.01
to 10.
Query integration rate.
Path to configure voltage compliance:
Specify voltage limit for I-Source; -210 to
210.
Query voltage compliance limit.
In voltage compliance?: 1 (yes), 0 (no).
Path to configure resistance:
Select ohms mode (MANual or AUTO).
Query ohms mode.
Enable or disable offset-compensated ohms.
Query state of offset-compensated ohms.
SCPI Command Reference
Default
parameter SCPI
10
1.05e-4
21
ON
0.21
10
21
MANual
OFF
17-11
Source
memory
1
1
2
2
2
2

Advertisement

Table of Contents
loading

Table of Contents