E-4
IEEE-488 and SCPI Conformance Information
Table E-2
Coupled commands
Command
:SENSe...:RANGe:UPPER
:SENSe...:NPLC
:SOURce...:RANGe
:SOURce...:STARt
:SOURce...:STOP
:SOURce...:STEP
:SOURce...:POINts
:SOURce...:CENTer
:SOURce...:SPAN
REN, GTL
:SYSTem:MEMory:INITialize
Also changes
:SENSe...:RANGe:AUTO
:NPLC for all other functions
:SOURce...:RANGe:AUTO
:SOURce...:STEP
:SOURce...:CENTer
:SOURce...:SPAN
:SOURce...:STEP
:SOURce...:CENTer
:SOURce...:SPAN
:SOURce...:POINts
:SOURce...:STEP
:SOURce...:STARt
:SOURce...:STOP
:SOURce...:STEP
:SOURce...:STARt
:SOURce...:STOP
:SOURce...:STEP
See local and remote transition in Section 13,
Differences: remote vs. local operation.
See command description in Section 17, SYSTem
subsystem.