Set Scaling Factor - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

:POINts <NRf>
:SOURCe:MEMory:POINts <NRf>
Parameters
Description
:STARt <NRf>
:SOURCe:MEMory:STARt <NRf>
Parameters
Description
:RECall <NRf>
:SOURCe:MEMory:RECall <NRf>
Parameters
Description

Set scaling factor

:TRIGgered:SFACtor <n>
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor <n>
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor <n>
Parameters
Query
Description
NOTE
These commands work only with source memory sweeps.
<NRf> = 1 to 100
This command is used to specify the number of points for the sweep. For
example, if you saved setups in memory locations 1 through 12 for a sweep,
specify a 12-point sweep using this command.
<NRf> = 1 to 100
This command is used to set the starting location of a Source Memory
Sweep. For example, for setups saved in memory locations 98 through 5,
specify a starting location of 98.
<NRf> = 1 to 100
This command is used to return the SourceMeter to the setup stored at the
specified memory location.
<n> =
-999.9999e+18 to 999.9999e+18
:SFACTor?
:SFAC instructs the SourceMeter to source the scaling factor times the pre-
vious source memory location value. For example, if 10.0V is stored in the
first source memory (Source I, Measure V Mode), and the unit is in the
Source V, Measure I mode with :SFAC set to 0.1 and enabled, the unit will
output 1.0V for the second source memory location.
SCPI Command Reference
Specify number of sweep points to exe-
cute
Number of sweep points
Select Source Memory Sweep start
location
Specify memory location
Return to specified setup
Specify memory location
Set current scaling factor
Set voltage scaling factor
Scaling factor
Query scaling factor
17-79