Source Delay - Keithley 6430 Instruction Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

Front panel V-source protection
To program V-source protection from the front panel:
1.
Press CONFIG then SOURCE V.
2.
Select PROTECTION from the displayed choices, then press ENTER.
3.
Select the desired protection value, then press ENTER.
4.
Press EXIT to return to normal display.
Remote command V-source protection
Use the :SOURce:VOLTage:PROTection command to program the V-source protection
value via remote. See Section 17 for details. For example, the following command sets the pro-
tection value to 20V:
:SOUR:VOLT:PROT 20

Source delay

The source delay options are used to set the settling time for the source. This source delay is
the delay phase of the Source-delay-measure cycle . See Section 5. The auto delay period is cur-
rent range dependent (Table 3-4).
T a ble 3-4
Auto source delay
SourceMeter
I-Range
Auto Delay
1pA
10pA
100pA
1nA
10nA
100nA
1µA
10µA
100µA
1mA
10mA
100mA
Manual delay can be set from 0000.00000 to 9999.99800 seconds. Manually setting the
delay disables auto delay.
1sec
350msec
50msec
35msec
10msec
6msec
5msec
5msec
3msec
3msec
3msec
3msec
Basic Source-Measure Operation