Keithley 6430 Instruction Manual page 192

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

9-16
Sweep Operation
Step 5: Run sweep.
To run the sweep, press the SWEEP key. After the sweep is completed, turn the output off by
pressing the ON/OFF OUTPUT key.
Step 6: Read buffer.
Use the RECALL key to access the source-measure readings stored in the buffer. Use the
TOGGLE to display statistical information.
Performing a custom sweep
Step 1: Configure source-measure functions.
Configure the SourceMeter for the desired source-measure operations as follows:
1.
2.
3.
Step 2: Configure sweep.
Configure the sweep as follows:
1.
2.
3.
4.
5.
6.
7.
Step 3: Set delay.
Set the source delay as follows:
1.
2.
3.
4.
Select the desired source function by pressing SOURCE V or SOURCE I.
Set the source level and compliance limit to the desired values.
Press MEAS V or MEAS I to select the desired measurement function, then choose the
desired measurement range.
Press CONFIG then SWEEP.
Select TYPE, then press ENTER.
Select CUSTOM, then press ENTER.
Use the displayed menu selections to enter the desired # POINTS, individual point val-
ues (ADJUST POINTS), and INIT (initial) value.
From the CONFIGURE SWEEPS menu, select SWEEP COUNT, press ENTER, then
choose FINITE or INFINITE as desired.
Again from the CONFIGURE SWEEPS menu, choose SOURCE RANGING, press
ENTER, then select BEST FIXED, AUTO RANGE, or FIXED as appropriate.
Press EXIT to return to normal display.
Press CONFIG then SOURCE V or SOURCE I depending on the selected source
function.
Select DELAY, then press ENTER.
Set the delay to the desired value, then press ENTER.
Press EXIT to return to normal display.