18-2
Performance Verification

Introduction

Use the procedures in this section to verify that Model 6430 accuracy is within the limits
stated in the instrument's one-year accuracy specifications. You can perform these verification
procedures:
WARNING
NOTE
Verification test requirements
Be sure that you perform the verification tests:
Environmental conditions
Conduct your performance verification procedures in a test environment with:
When you first receive the instrument to make sure that it was not damaged during
shipment.
To verify that the unit meets factory specifications.
To determine if calibration is required.
Following calibration to make sure it was performed properly.
The information in this section is intended for qualified service personnel
only. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, which
could cause personal injury or death if contacted. Use standard safety pre-
cautions when working with hazardous voltages.
The Remote PreAmp connectors carry hazardous voltage. To prevent risk
of electric shock, connectors must be fully mated or safety covers must be
placed over the open connectors. Proper installation requires that the oper-
ator is protected from exposed voltages by insulation or barriers.
If the instrument is still under warranty and its performance is outside specified lim-
its, contact your Keithley representative or the factory to determine the correct
course of action.
Under the proper environmental conditions.
After the specified warm-up period.
Using the correct line voltage.
Using the proper test equipment.
Using the specified output signals and reading limits.
An ambient temperature of 18-28°C (65-82°F).
A relative humidity of less than 60% unless otherwise noted.