17-104 SCPI Command Reference
:DELay <n>
:TRIGger[:SEQuence[1]]:DELay <n>
Parameters
Query
Description
:SOURce <name>
:ARM[:SEQuence[1]][LAYer[1]]:SOURce <name>
:TRIGger[:SEQuence[1]]:SOURce <name>
Parameters
NOTE
Query
Description
NOTE
<n> =
0 to 999.9999
DEFault
MINimum
MAXimum
:DELay?
:DELay? DEFault
:DELay? MINimum
:DELay? MAXimum
The delay is used to delay operation in the trigger layer. After the pro-
grammed trigger event occurs, the instrument waits until the delay period
expires before performing the Device Action.
<name> =
IMMediate
TLINk
TIMer
MANual
BUS
NSTest
PSTest
BSTest
Only IMMediate and TLINk are available as trigger layer control sources.
:SOURce?
These commands are used to select the event control source. With IMMedi-
ate selected, operation immediately continues.
A specific event can be used to control operation. With TLINk selected,
operation continues when a trigger pulse is received via the Trigger Link.
The following control sources are not available for the trigger layer.
With TIMer selected, the event occurs at the beginning of the timer interval,
and every time it times out. For example, if the timer is programmed for a
30 second interval, the first pass through the control source occurs immedi-
ately. Subsequent arm events will then occur every 30 seconds. The interval
for the timer is set using the :TIMer command.
With MANual selected, the event occurs when the TRIG key is pressed.
Set trigger layer delay
Specify delay in seconds
0 second delay
0 second delay
999.9999 second delay
Query the programmed delay
Query the *RST default delay
Query the lowest allowable delay
Query the largest allowable delay
Specify arm event control source
Specify trigger event control source
Pass operation through immediately
Select Trigger Link trigger as event
Select timer as event
Select manual event
Select bus trigger as event
Select low SOT pulse as event
Select high SOT pulse as event
Select high or low SOT pulse as event
Query programmed control source