You have completed the scale calibration procedure. Before replacing the Calibration Switch
Figure 51. Calibration Cover and Switch Location
A
B
Figure 52. If one is present, move EAS antenna to access the Calibration Switch
1
Product Reference Guide
Cover and sealing the scale, you must now continue with the calibration verification tests
to complete the scale's calibration. If you are weighing in pounds, continue with the
following procedures to verify that the scale was successfully calibrated and that it weighs
properly. If you are weighing in kilograms, turn to the instructions titled, "Calibration
Verification (Kilograms)" on page 85.
A
Calibration Seal Location
Calibration Access Cover
If an EAS antenna is present, it must be lifted in order to access the calibration
Figure
switch. See
NOTE
1
2
EAS Antenna
B
Calibration Switch Opening. Use a
long, blunt object to reach in and
C
press the Calibration Switch.
52.
Calibration Switch Cover
2
C
79