Site Preparation and Installation
Figure 34. Using the Lift Handles
8. Lower the unit into the counter opening, ensuring that none of the cables are pinched,
pierced or crimped.
9. View the bubble level (if present
scanner is level. As shown in the insert of
within the circle indicated. The air bubble can touch the circle, but must not overlap it.
Apply shims beneath the scanner flange or adjust the leveling feet to allow level installa-
tion.
10. Reinstall the All Weighs™ Platter and verify that it is flush or just below flush with the
countertop. This is necessary to provide smooth scanning from either direction. Make
adjustments as needed to align the platter with the counter by moving support rails up or
down, or consider installing screws in positions that will allow their use in adjusting the
unit's position. Ensure that leveling is maintained if adjustments are made.
NOTE
11. Once installation is complete, proceed with the
follow.
1. The bubble level is present only on certain models.
48
Bubble Level
1
) located on the scanner's spider assembly to ensure the
Failure to install a scanning-scale in a stable and level position will inhibit weigh-
ing, calibration and zeroing operations. The platter MUST make unobstructed
contact with all of its supports for proper weighing operation.