Dilution Or Sample Unit Incompatible; Direction - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Dilution or
sample unit
incompatible

Direction

Page 94
Manual reprocessing error: The method used for reprocessing has
a sample definition different from the curve generated.
Example: you run a titration with a sample dilution programmed and
try to reprocess this curve in manual mode with no dilution
programmed for the sample.
Edit the Sample parameters of the method to be reprocessed.
Refer to "Reprocessing a curve (Manual mode)", page
Determines whether the addition of titrant increases or decreases the
pH or mV value.
This parameter informs the Titration Manager if the end point or the
Stop point is already passed before the start of the titration.
Example:
If the titration is performed with an acidic titrant on an alkaline sample,
the addition of titrant will decrease the pH value. In this case, select
Direction = Decreasing pH.
Enter in:
Edit method > Parameters menu
Edit reagent > Calibration parameters menu
TitraLab 870 Reference Manual
203.