Qc Data Menu; Qc Id; Qc Not Required; Qc Periodicity Elapsed - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

QC data menu

QC ID

QC not required

QC periodicity
elapsed
This menu is available for measurement and titration methods.
To access:
1. Enter the Main window.
2. Select or create a method.
3. Select QC sample = Yes in the Edit method menu.
4. Use the LEFT/RIGHT arrow keys to move to the last Edit method
display.
5. Press 5 QC data.
Enter the name of the QC sample (16 alphanumeric characters). This
name is called up when you run a QC sample analysis.
Message appears at the start of a sequence, that originally included a
method programmed with a QC sample.
The method to has now been reprogrammed without QC sample.
Go to Sequence/Sample stack, Edit sequence screen and remove the
QC sample from the sequence.
This message is displayed at the start of a method requiring a QC
sample. The Periodicity, entered in the QC Data screen of the method
has elapsed. For example, if Periodicity = 10 samples, then a QC
sample must be performed every 10 samples.
Press ✓ and run a QC analysis.
TitraLab 870 Reference Manual
Page 183