Create Method; Create Reagent; Current Value - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®

Create method

Create reagent

Current value

Page 88
1. Enter the Main window.
2. Select Working mode = Method.
3. Press 4 then 1.
4. Press ✓ in the New method screen.
5. Enter a method name (up to 17 alphanumerical characters).
6. Press 1 to create the method.
Go to Edit method screen and enter the method parameters.
Refer to "Programming method", page
1. Enter the Reagents window.
2. Press 4 then 1.
3. Press ✓ in the New reagent menu.
4. Select From = Other.
The option From = Catalogue allows you to create a reagent
from a list of commonly used reagents.
5. In the ID field, enter the reagent name (up to 14 alphanumeric
characters). It is advised to enter the chemical formula of the
reagent followed by its concentration (e.g. HCl 0.1).
6. Enter the "approximate" titre of the reagent (5 characters) in the
Target titre field.
7. Enter the unit indicated on the reagent bottle (mM = mmol/l, M =
mol/l, mN = meq/l or N = eq/l).
8. Press 1 twice to create the reagent.
9. Enter the reagent parameters.
This is the current sent to the Pt-Pt socket on the titration system
(Titration Manager or ABU52). This parameter is available if
mV(i>0) has been selected for Measurement in the Edit
method or Edit reagent menus.
Enter in:
Method parameters menu
Reagent parameters menu
Range available:
-1000 to +1000 µA in steps of 1 µA
TitraLab 870 Reference Manual
180.