Addition
method -
programmation
For a 1-addition method
Refer to "Addition method - definition", page
For a 2 or 3-addition method
At the Simultaneous additions, select whether you want
to perform the additions simultaneously or one after the other.
• In the case of simultaneous additions:
Enter the delay to wait at the end of the 2 or 3 additions (00:00 to 99:59 min:s).
Enter for each addition, the reagent name and the volume to be added (1 µl to
999 ml).
• In the case of additions performed one after the other:
Enter for each addition, the reagent name, the volume of reagent to be added
(1 µl to 999 ml) and the delay to wait at the end of the addition (00:00 to
99:59 min:s).
For the Auxiliary output parameter,
In an Addition method, it is not possible to run 2 additions of
the same reagent.
In an titration method, it is also possible to run one reagent
addition before the titration starts,
TitraLab 870 Reference Manual
48.
see "Auxiliary output", page
see "Addition", page
55.
47.