TitraLab
®
Running a
manual dosing
To prepare a manual dosing,
Automatic add. = No
An increment is added each time key 1 is pressed.
Automatic add. = Yes
The reagent is added automatically if you press 1. Dosing stops when
the Max. volume has been delivered or if you press 1 again.
If no measurements: the volume
already dispensed is displayed.
Curve volume (in ml) = f(time)
is displayed live.
Displaying the direct or derivative curve: press Del.
Condition: Measurements must have been performed
Stopping a manual dosing: press Stop.
The Reagent window is displayed.
TitraLab 870 Reference Manual
see "Manual dosing", page
Adjust the increment size (0.01 ml to 10 ml).
Select Yes to add the increments continuously
Select No to add a given number of
increments.
Press 1 to start dosing the reagent
(one increment is added or reagent is added
automatically increment by increment ).
If measurements have been selected, pH
(or mV) and/or temperature are displayed live.
With measurements: the measured
signal and/or temperature are also
displayed. Curve measured signal
= f(volume) is displayed live.
Select the curve to display.
Press 1 to quit the manual dosing.
Use the RIGHT/LEFT arrow keys to move
along the curve and display the point
coordinates shown by the cursor.
148.