TitraLab
®

Programming SAC sequences

A SAC sequence is a sequence of methods with automatic change of the beakers. A sample
changer (SAC80, SAC90, SAC850 or SAC950) is used.
1.
Press Stop for 3 seconds to
enter the Setup menu.
4.
Select SAC Sequence.
2.
Press 1.
5
Press 2.
TitraLab 870 Reference Manual
3.
Select a Sample Changer
(SAC80, SAC90, SAC850 or
SAC950). Enter the parameters
of the sample changer selected
(number of rinses, rinse time,
etc.).
Press Esc then 5 (Exit) to quit
the Setup menu.
6.
Enter a name for the sequence.
Page 35