Method Results Menu; Method Type Incompatible For A Reprocessing; Method Wrong Type; Min. Cell Cst - Max. Cell Cst - Radiometer Analytical TitraLab 870 Reference Manual

Ph/ep/ip/ec/ise titration workstation
Table of Contents

Advertisement

TitraLab
®
Method results
menu
Method type
incompatible for
a reprocessing
Method wrong
type
Min. cell cst -
Max. cell cst
Page 154
To access:
1. Press 4 in the Main window.
2. Press 2 Edit method.
3. Use arrow keys to move to last display.
4. Press 3 Results.
Manual reprocessing error: The method selected in reprocessing
mode is not correct (wrong method mode). Example: you have
selected a Measurement or an Addition method and try to run it in
reprocessing mode.
Reprocessing with a Measurement or Addition method is not possible.
Select another method.
Refer to "Reprocessing a curve (Manual mode)", page
This message appears at the start of a coupled method if it contains a
Coupled method.
A Coupled method cannot be part of a Coupled method.
Acceptance range for the cell constant.
If the cell constant lies outside the defined range, the conductivity cell
calibration must be repeated.
Enter in:
Edit electrode > Result menu (Conductivity type electrodes)
Range available:
Min. cell cst = 0.05 cm
Max. cell cst = Min. cell cst to 15.00 cm
TitraLab 870 Reference Manual
-1
to Max. cell cst
-1
203.